AOIP CALYS 150 Advanced Documenting Multifunction Calibrator Hermometer

Features

  • Scaling allows process signals to be displayed in % of FS or in all other units. This function also allows sensors to be corrected after a calibration.
  • With 100 values manually set, CALYS 150 enables users to draw a generation curve.
  • CALYS 150 is able to be used as a transmitter. Measurement input is copied on the output with scaling.
  • In temperature or pressure mode, CALYS 150 can control electronic thermostat and press stat trigger levels.
Enquiry
Category:

AOIP CALYS 150 field documenting multi function calibrator. . It’s the ideal tool for advanced maintenance and testing across all industries. Suitable for all field and lab measurements, it can simultaneously measure and generate over two isolated channels various signals of temperature, resistance, process, pressure and frequency in one single instrument.

AOIP CALYS 150 not only simulates (IN / OUT) but also acts as a dual-channel thermometer (IN / IN) for comparison calibration. It calibrates HART transmitters (HART communicator integrated into ACL500 modem) and thermistors. Providing extended functionalities (temperature simulation, scaling, steps, synthesizer, statistical functions…) and audit trails, CALYS 150 complies with both 21 CFR Part 11 and NADCAP Heat Treatment standards and makes advanced data exploitation and full data traceability easier.

Performance Specifications:

  • Temperature: Accurate measurement and simulation up to 0.005 % RDG
  • Resistance: Delivers precision up to 0.006 % RDG and 50 KΩ range
  • Current: Offers up to 0.007 % RDG and 100 mA range + Loop Supply 24 V
  • Voltage: Provides up to 0.005 % RDG and 50 V range
  • Frequency: Achieves up to 0.01 % RDG and 100 KHz range
  • Pressure: Compatible with an external pressure module for comparison calibration using a pressure pump.

With this easy-to-use tool, calibration tasks can be swiftly completed across the entire process chain. Take the 900 g documenting process calibrator to the field with you during the whole week with 10 calibration procedures stored in the device. Run the procedure after connecting the probes to the instrument and save the results for onsite easy and quick calibration. Back at the office, upload data to a computer to generate customized calibration certificates using dedicated DATACAL software.

With an IP 54 rating and a shock-resistant rubber holster, CALYS 150 is tough and durable. It features “easyconnect” terminals and a wide backlit display, ensuring ease of use even in harsh or dark conditions. When paired with an external pressure module (ref. ACL433), CALYS 150 can measure and simulate pressure using a pressure pump for comparison calibration. It also has the ability to control baths and dry-blocks with the appropriate cable (ref. ACL600).

Easy connection system

Connect your probes easily by pressing on the terminal top and inserting wires up to 3 mm or 10 AWG in diameter, along with compensated thermocouple connectors. Wires are securely held between two brass plates, providing thermal stability and excellent cold junction compensation for thermocouples. This system also allows for connecting 4 mm banana plugs and security connectors on the terminal top.

Specification

Size 210 x 110 x 50 mm
Weight 900 g
Display 240 x 320 pixel liquid crystal graphical display with backlite and contrast control Display of result as table of values or trend curve
Power supply 230 V ± 10 %, 50/60 Hz
Battery Type:lithium-ion Charging Time:  3 hours lifetime: 8hours min
Communication ports USB
Storage capacity Up to 10 full configurations (Input / output type, range…) 10,000 data into one or several measurement campaigns, i.e. more than one week work with configurations, measurements, calibration procedures and reports
Reference range 23°C ± 5°C (RH: 45 to 75 % w/o condensing)
Operating reference range -10 to 50°C (RH: 20 to 80 % w/o condensing)
Limit operating range -15°C to +55°C (RH: 10 to 80 % w/o condensing) (70% at 55°C)
Storage temperature limits -30°C to +60°C
Maximum height 0 to 2000 m
IP protection IP54 according to EN60529
Protections • Electronic protection up to 250 V for ‘voltage’ wires
• Fuse protection for ‘current’ wires
• Protection against ‘current’ circuit breaking during inductive resistance measurements
Class In accordance with EN 61010-1 Category II, pollution 2
Rated voltage 60 V
Chocks and vibrations EN 61010-1
EMC conformity Immunity:
• EN 61000-4-2
• EN 61000-4-3
• EN 61000-4-5
• EN 61000-4-6
• EN 61000-4-11EN 61000-4-4 Conducted and radiated emissions:
• EN 55022, class B
• EN 61000-3-2
• EN 61000-3-3

Download

AOIP CALYS 150 Advanced documenting multifunction calibrator thermometer-DpstarGroup