Lutron CL-2006 – Chlorine Meter

Features

  • Free chlorine, total Chlorine measurement.
  • Range : 0 to 3.5 mg/L (ppm).
  • Range : Auto range.
  • Measuring principal : Photo meter method.
  • Light source : LED, 525 nm.
  • Record Max, Min. value.
  • 10 PCs Free and Total powder DPD are included.
  • Free and total 1.0 mg/L standard solution are the
    standard accessories.
  • Power : 1.5 V battery (UM-4/AAA) x 6 PCs.
  • The measuring method is an adaptation of the USEPA Method 330.5 and Standard Method.
Enquiry

Lutron CL-2006 Chlorine Meter offers measurement capabilities for both free and total chlorine levels within a range of 0 to 3.5 mg/L (ppm). It employs an auto-ranging feature and utilizes a photometer method with an LED light source at 525 nm. The device records maximum and minimum values, and it comes with 10 packets of Free and Total powder DPD for testing. Additionally, it includes 1.0 mg/L standard solutions for both free and total chlorine measurements. Power is provided by six 1.5V batteries (UM-4/AAA). The measuring method is based on adaptations of the USEPA Method 330.5 and Standard Method protocols.

Specification

Circuit Custom one-chip of  microprocessor  LSI circuit.
Display LCD size : 41 mm x 34 mm
Range • Free chlorine ( CL ) : 0.00 to 3.50 ppm ( mg/L )
• Total chlorine ( CL ) : 0.00 to 3.50 ppm ( mg/L ).
Resolution 0.01 ppm ( mg/L ).
Accuracy ± 0.02 ppm ( mg/L ).
@ 1.00 ppm ( mg/L )
Light source LED, 525 nm
Light detector Photo diode
Response time Less than 10 seconds.
Sample volume 10 mL
Data Hold Freeze the display reading.
Memory Recall Maximum & Minimum value.
Method The measuring method is an adapta-tion of the USEPA Method 330.5 and Standard Method 4500-Cl G.

•  The reaction between free ( total ) hlorine and the DPD reagent cause a pink tint in the sample.

Display Sampling Time Approx. 1 second.
Power off Auto shut off saves battery life or manual off by push button
Calibration points Zero chlorine.
• 1.0 ppm ( Free chlorine )
• 1.00 ppm ( Total chlorine ).
Operating Temperature 0 to 50
Operating Humidity Less than 85% R.H.
Power Supply DC 1.5 V battery ( UM4, AAA ) x 6 PCs, or equivalent
Power Current • Stand by Approx. DC 4 mA
• Testing Approx. DC 12 mA.
Weight 320 g/0.70 LB. @ Battery is included.
Dimension 155 x 76 x 62 mm ( 6.1 x 3.0 x 2.4 inch)
Accessories Included • Hard carrying case, CA-08………………………………………………………….. 1 PC
• Instruction manual………………………………………………………………………… 1 PC
• 1.0 ppm Total Chlorine standard solution, CT-01………………………….. 1 PC
• Zero Chlorine standard solution, CL-01………………………………………… 1 PC
• Empty testing bottle……………………………………………………………………….2 PCs
• Clean cloth…………………………………………………………………………………….1 PC
• Free Chlorine DPD powder………………………………………………………….10 PCs
• Total Chlorine DPD powder………………………………………………………….10 PCs
Optional Accessories • Free Chlorine DPD powder ( 10 PCs ),
Model : CFP-10• Total Chlorine DPD powder ( 10 PCs ),
Model : CTP-10• Empty testing bottle,
Model : 0601• 1.0 ppm Free Chlorine standard solution,
Model : CF-01• 1.0 ppm Total Chlorine standard solution,
Model : CT-01

• Zero Chlorine standard solution
Model : CL-01

Download