Handheld Thermometer For Thermocouples And Or Resistive Probes With Memory – AOIP TM TM6602 / TM6612 / TM6630

Features

  • High accuracy: 0.02% reading with an adjustable resolution of 1 µV (TM 6602 / 6630) and 1 mΩ (TM 6612 / 6630)
  • Values displayed in °C, °F, mV and Ω
  • Low temperature coefficient: 10 ppm /°C for thermocouples and 7 ppm/°C for resistance
  • Measurements with HOLD function
  • Correction of calibrated sensors
  • Display of minimum, maximum and average value
  • Backlight
  • 10,000 values stored and displayed graphically
Enquiry

Easy-to-use and equipped with a large graphical display, TM thermometers are the perfect field tools for daily maintenance and commissioning of temperature probes. 3 models are available with 0.02% accuracy as follows:

• TM 6602 for thermocouple measurement
• TM 6612 for RTD measurement
• TM 6630 for both thermocouple and RTD measurement

With a very low temperature coefficient (10 ppm/°C in thermocouples and 7ppm/°C in resistance), IP54 protection and robust design, TM thermometers are suitable for onsite use even in demanding environmental conditions. They are widely used in the energy field, engineering sector, metal industry and automotive field.

TM thermometers use a graphical interface making programming and reading easier, under graphical or test format. Due to memory feature (10,000 values) and extended functionalities (corrections of calibrated sensors, statistical functions…), TM are well adapted to different process job procedures and ensure full data traceability as well as advanced data exploitation. Use them together with DATACAL calibration software to display, manage data and issue your own customized calibration certificates. TM thermometers are delivered with 4 alkaline AA batteries in standard. The optional battery charger used a rechargeable battery.

Specification

Size 157 x 85 x 45 mm
Weight 306 g
Display 160 x 160 pixel liquid crystal graphical display with backlite
Display of result as table of values or trend curve
Power supply 4 AA batteries 1.5 V or rechargeable Ni-Mh batteries with internal charger in option
Communication ports USB
Storage capacity 10,000 data with date and time into one or several acquisition bursts
Reference range 23°C ± 5°C (RH: 45 to 75 % w/o condensing)
Operating reference range -10 to 50°C (RH: 20 to 80 % w/o condensing)
Limit operating range -15°C to +55°C (RH: 10 to 80 % w/o condensing) (70% at 55°C)
Storage temperature limits -30°C to +60°C
Maximum height 0 to 2,200 m
IP protection IP54 according to EN60529
Protections • Electronic protection up to 250 V for ‘voltage’ wires
• Fuse protection for ‘current’ wires
• Protection against ‘current’ circuit breaking during inductive resistance measurements
Class In accordance with EN 61010-1
Category II, pollution 2
Rated voltage 60 V
Chocks and vibrations EN 61010-
EMC conformity Immunity:
• EN 61000-4-2
• EN 61000-4-3
• EN 61000-4-5
• EN 61000-4-6
• EN 61000-4-11EN 61000-4-4
Conducted and radiated emissions:
• EN 55022, class B
• EN 61000-3-2
• EN 61000-3-3

Download