Temperature Converter – Shimaden CP2RT Series

CP2RT functions to convert thermocouple and R.T.D. input signals into DC signals. Standard feature: Isolation, Linear output and Burnout function 6 toxic substances, which are subject to RoHS regulations, are contained. However, the amount of toxic substances contained does not exceed standardized values.

Enquiry

CP2RT functions to convert thermocouple and R.T.D. input signals into DC signals. Standard feature: Isolation, Linear output and Burnout function 6 toxic substances, which are subject to RoHS regulations, are contained. However, the amount of toxic substances contained does not exceed standardized values.

Specification

Thermocouple Input T, E, J, K, N, R, S, B
Conversion accuracy ±0.3% FS / Linearized output (at 23 °C)
Influence of ambient temperature ±0.2% FS for temperature change of 10 °C
Measurement range Refer to “Ordering Information”.
Conversion output Refer to “Measuring Range Codes”
Cold Junction Temp. Compensation Range 5–45°C
Input Resistance 1MΩ
External Resistance Tolerance 100 Ω max
R.T.D. Input Pt100 / JPt100 (3-wire type)
Conversion Accuracy ±0.3% FS / Linearized output (at 23 °C)
Influence of ambient temperature ±0.2% FS for temperature change of 10 °C
Measurement range Refer to “Measuring Range Codes”
Conversion output Refer to “Ordering Information”.
Amperage 1mA
R.T.D. Lead Wire Tolerable Resistance 5Ω max. per wire
Response speed 500ms max. (0–90%)
Reproducibility ± 0.1% FS
Burnout Circuit Standard (upscale)
Conversion Output Adjustment Range ±4% (ZREO and SPAN)

Ordering Information

ITEMS CODE SPECIFICATIONS
SERIES CP2RT- Temperature Converter
INPUT 1 Thermocouple, T, E, J, K, N, R, S, B Input Resistance:1MΩ
2 R.T.D. (Pt100/JPt100), 3-wire type Amperage: 1mA
CONVERSION OUTPUT 1 0–10mV DC / FS Output Resistance: 5 Ω
2 0–100mV DC / FS Output Resistance: 50 Ω
3 0–1V DC / FS Max. Current: 2mA
4 0–10V DC / FS Max. Current: 2mA
5 1–5V DC / FS Max. Current: 2mA
6 4–20mA DC / FS Load Resistance: 600Ω max
9 Others (Please consult before ordering.)
POWER SUPPLY 02- 24 V DC±10%
13- 100–110 V AC±10% 50/60Hz
14- 110–120 V AC±10% 50/60Hz
15- 200–220 V AC±10% 50/60Hz
16- 220–240 V AC±10% 50/60Hz
99- Others (Please consult before ordering.)
INPUT STANDARD F JIS PT100
J JIS JPT100 and Thermocouple
D DIN
A ANSI
X Others (Please consult before ordering.)
TYPE OF INPUT T Thermocouple T Cold Junction Temperature Compensator
(Externally installed)
E Thermocouple E
J Thermocouple J
K Thermocouple K
H Thermocouple N
R Thermocouple R
S Thermocouple S
B Thermocouple B Cold Junction Temperature Compensator (Not Available)
P R. T. D. Pt100
X Others (Please consult before ordering.)
MEASURING RANGE °°° See Measuring Range Codes.
LEGEND C °C
F °F
X Others (Please consult before ordering.)
REMARKS 0 Without
9 With (Please consult before ordering.)

Download