Beamex MC6-T Multifunction Temperature Calibrator & Communicator

Features

  • Superior Metrological Performance: Provides high accuracy with active multizone temperature control technology, ensuring an excellent temperature gradient.
  • Stable and Safe: Advanced methods ensure stability and accuracy, even in fluctuating environmental conditions.
  • Advanced Functionality: Calibrates temperature, pressure, and electrical signals with full automation capability. Works with a rechargeable battery for field use.
  • Field Communicator: Supports HART, FOUNDATION Fieldbus, and Profibus PA protocols for field instrument calibration.
  • Digital Data Flow: Fully digitalized calibration process when connected to Beamex calibration management software, ensuring secure data flow and traceability.
Enquiry

Beamex MC6-T combines a state-of-the-art temperature dry block with a built-in multifunction process calibrator and communicator. It offers unmatched versatility by providing calibration capabilities for temperature, pressure, and electrical signals, as well as acting as a multi-bus communicator for fieldbus instruments. Designed for field use, the MC6-T is built to handle varying environmental conditions while maintaining accuracy. Its patented temperature control algorithm ensures fast, efficient heating and cooling without overshoots, improving calibration efficiency.

Benefits

  • Versatile Calibration: Capable of temperature, pressure, and electrical calibration in one device.
  • Fast and Efficient: Patented technology provides quick heating and cooling, saving time during calibrations.
  • Advanced Safety Features: Includes tilt sensor, warning light, and overheat protection for safe operation.
  • Automated Documentation: The MC6-T automates calibration tasks and documents data digitally, reducing manual input and improving data accuracy.

Capabilities

  • Temperature Generation
  • Pressure Measurement: Internal barometric and external pressure modules
  • Voltage Measurement: ±1 V and −1…60 VDC
  • Current Measurement: ±100 mA (internal or external supply)
  • Frequency Measurement: 0…50 kHz
  • Pulse Counting: 0…10 Mpulse
  • Switch State Sensing: Dry/wet switch
  • Built-in 24 VDC Loop Supply: HART impedance or FF/PA impedance
  • Voltage Generation: ±1 V and −3…24 VDC
  • Current Generation: 0…55 mA (active/passive)
  • Resistance Measurement and Simulation: Two simultaneous channels (0…4 kΩ)
  • RTD Measurement and Simulation: Two simultaneous channels
  • TC Measurement and Simulation: Two simultaneous channels (universal connector/mini-plug)
  • Frequency Generation: 0…50 kHz
  • Pulse Queue Generation: 0…10 Mpulse
  • HART, FOUNDATION Fieldbus, and Profibus PA Communication

Specification

Dimensions 322 mm x 180 mm x 298 mm (12,68″ x 7,09″ x 11,73″)
Weight MC6-T150: 9.4 kg (20.7 lbs)
MC6-T660: 8.6 kg (18.96 lbs)
Display 5.7” Diagonal 640 x 480 TFT LCD Module
Touch Panel 5-wire resistive touch screen
Keyboard Membrane keyboard
Backlight LED backlight, adjustable brightness
Power requirements 230 V ±10%, 50/60 Hz, 380 W (MC6-T150, 1560 W (MC6-T660)
115 V ±10%, 50/60 Hz, 380 W (MC6-T150), 1560 W (MC6-T660)
Fuse size (MC6-T660) 230 V: T 8A 250V / 115 V: T 16A 250V
Fuse size (MC6-T150) 230 V: T 3.15A 250V / 115 V: T 3.15A 250V
Max. input voltage 30 V AC, 60 V DC
Operating temperature 0 … 45 °C (32 … 113 °F)
Operating humidity 0 … 90% R.H. non condensing
Storage temperature -20 … 60 °C (-4 … 140 °F)
Computer interface USB
Calibration certificate Accredited calibration certificate
Warmup time Specifications valid after a 5 minute warmup period
Battery type Rechargeable lithium-ion polymer, 4300 mAh, 11.1 V
Charging time Approximately 4 hours
Battery operation time 10 … 16 hours
Battery operated functions All functions except temperature control and R3 measurement
Safety Directive 2014/35/EU, EN 61010-1:2010 + A1:2019 + AC:2019-04
EMC Directive 2014/30/EU, EN IEC 61326-1:2021
RoHS compliance RoHS II Directive 2011/65/EU, EN IEC 63000:2018
Drop EN 61010-1:2013
Warranty Warranty 3 years, 1 year for battery pack.

 

Temperature Specification

Feature MC6-T150 MC6-T660
Temperature range at 23 °C (73 °F) -30 … 150 °C (-22 … 302 °F) 50 … 660 °C (122 … 1220 °F)
±0.2 °C at 50 °C
Display uncertainty with internal reference 1) ±0.15 °C ±0.3 °C at 420 °C
±0.5 °C at 660 °C
±0.02 °C at 50 °C
Stability 2) ±0.01 °C ±0.03 °C at 420 °C
±0.04 °C at 660 °C
±0.05 °C at 50 °C
Axial uniformity at 40 mm ±0.05 °C ±0.25 °C at 420 °C
±0.40 °C at 660 °C
±0.10 °C at 50 °C
Axial uniformity at 60 mm ±0.07 °C ±0.40 °C at 420 °C
±0.60 °C at 660 °C
Radial uniformity Difference between borings ±0.01 °C at 50 °C
±0.01 °C ±0.05 °C at 420 °C
±0.08 °C at 660 °C
Loading effect with internal reference sensor With 4 pcs 6 mm sensors ±0.02 °C at 50 °C
± 0.08 °C ±0.08 °C at 420 °C
±0.15 °C at 660 °C
Loading effect with external 6 mm reference sensor With 3 pcs 6 mm sensors ±0.01 °C at 50 °C
±0.005 °C ±0.02 °C at 420 °C
±0.03 °C at 660 °C
Hysteresis ±0.03 °C ±0.15 °C
Display resolution 0.001 °C /°F / K 0.001 °C /°F / K
Immersion depth 150 mm (5.9 in) 150 mm (5.9 in)
Insert outer dimensions 30 mm (1.18 in) 24.5 mm (0.96 in)
Heating time 23 to 150 °C:  19 min 50 to 660 °C:  15 min
-30 to 150 °C: 23 min
150 to 23 °C:  17 min 660 to 50 °C:  35 min
Cooling time 23 to -30 °C:  23 min 660 to 100 °C: 25 min
150 to -30 °C: 37 min
Stabilization time 3) 5 to 10 min 10 min

Download

Beamex MC6-T Temperature Calibrator & Communicator_DpstarGroup