Vaisala On-site Calibration Services For Continuous Monitoring Systems

Features

  • Achieve time and cost savings by preventing production downtime.
  • Ensure highly accurate measurements.
  • Provide certification and mandatory documentation for instrument calibration standards.
  • Free up staff to focus on primary work assignments.
  • Enjoy peace of mind with Vaisalacertified calibration expertise and components.
  • Take advantage of a wide array of available calibrations.
Enquiry

Vaisala’s on-site calibration for Continuous Monitoring Systems (CMS) helps boost reliability and profitability by reducing expensive equipment downtime. It eliminates the need for in-house or factory calibration, and allowing your staff to focus on what is truly important: your business.

Reliable On-site Calibration for Life Science Applications

Vaisala’s On-site Calibration Service offers many of benefits over the conventional options like sending equipment back to the manufacturer or doing calibration in-house. Reliable on-site calibration minimizes downtime by keeping your monitoring system in place and operational. If certain calibrations need the device to be removed from the process, Vaisala provides rental devices and swap services to keep monitoring uninterrupted and data consistent. By letting Vaisala handle calibration, you remove the need to invest in specialized calibration equipment and training. We offer various single and multi-point calibration options using specialized reference instruments and provide a certificate of NIST traceability. Plus, our system sends reminders when calibration is due to maintain accuracy and performance.

Complete Documentation

Vaisala On-site Calibration Service provides a thorough analysis with comprehensive paper and digital documentation, including a calibration certificate to ensure verification and standards compliance. Calibration reports, data sheets, and calibration labels can be supplied Vaisala on-site, with digital backup files saved to disk. Our highly accurate calibrations fulfill international standards and make it easy to comply with regulatory requirements.

Dpstar is a leading supplier of environmental measurement sensors for temperature, relative humidity, pressure, dew point, CO2, and other environmental condition monitoring for industrial process application and research. Since the ’90s, we earned a reputation as a trusted partner to our clients through our commitment and dedication. We are an authorized distributor and supplier of measuring sensors solutions from renowned manufacturers Vaisala. Vaisala is a global leader in environmental and industrial measurement. Building on 80 years of experience, Vaisala provides observations for a better world. Vaisala is a reliable partner for customers around the world, offering a comprehensive range of innovative observation and measurement products and services.

Specification

Single-point Calibration
Temperature
Range -90 … +70 °C
Unit Under Test acceptance limits ±1 °C
Calibration points One point at point of use
Adjustment Not available
Certificate Includes as-found/as-left data
Traceability NIST
Available for the following devices Vaisala Temperature Data Loggers
• DL1000
• DL1016
• DL1400
• DL1416
• DL2000
• HMT140
Reference instrument Vaisala Temperature Data Logger
Temperature for Liquid Nitrogen Applications
Range -196 °C
Unit Under Test acceptance limits ±3 °C
Calibration points One point at point of use
Adjustment Not available
Certificate Includes as-found/as-left data
Traceability NIST
Available for the following devices Vaisala Temperature Data Loggers DL1700
Reference instrument Vaisala Temperature Data Logger, Fluke 52 Series II
Relative Humidity
Range Ambient RH (within range of 10 … 90 %RH) at any temperature within range of +10 … +45 °C
Unit Under Test acceptance limits ±5 %RH
Calibration points One point at point of use
Adjustment Not available
Traceability NIST
Applicable loggers and transmitters Vaisala Humidity Data Logger DL2000, HMT140
Reference instrument Vaisala Humidity Data Logger
Carbon Dioxide
Range 0 … 20 % at point of use
Unit Under Test acceptance limits Application dependent
Calibration points One point at point of use
Adjustment Not available
Certificate Includes as-found/as-left data
Traceability NIST
Available for the following devices Vaisala Carbon Dioxide Transmitter Series GMT220
Reference instrument Vaisala GM70
Multi-point Calibration
Temperature
Range -90 … +90 °C
Unit Under Test acceptance limits ±0.5 °C
Calibration points Application-dependent
Adjustment Available on 3+ point calibrations
Certificate Includes as-found/as-left data and uncertainties
Traceability NIST
Available for the following devices Vaisala Temperature Data Loggers
with Probes
• DL1000
• DL1016
• DL1400
• DL1416
• HMT143
• HMT148
Reference instrument Ametek ETC159
Relative Humidity
Range 10 … 90 %RH at ambient temperature
Unit Under Test acceptance limits Loggers ±3 %RH
Calibration points Application-dependent
Adjustment Available on 3+ point calibrations
Certificate Includes as-found/as-left data and uncertainties
Traceability NIST
Available for the following devices Vaisala Relative Humidity Data Loggers HMT141, DL2000
Reference instrument Vaisala HMT330
Differential Pressure
Range -15 … 200 psi
Unit Under Test acceptance limits Range-dependent
Calibration points Application-dependent
Adjustment Available on 3+ point calibrations
Certificate Includes as-found/as-left data
Traceability NIST
Available for the following devices Vaisala Differential Pressure Transmitter Series PDT100
Reference instrument Fluke Process Calibrator / Pressure Modules
Current and Voltage
Range 0 … 5 VDC
0 … 10 VDC
0 … 20 mA
Unit Under Test acceptance limits ±0.15 %FS at +25 °C
Calibration points Match Factory Calibration
Adjustment Yes
Certificate Includes as-found/as-left data and uncertainties
Traceability NIST
Available for the following devices Vaisala Relative Data Loggers HMT140, DL4000, DL2000
Reference instrument National Instruments PHIe-4141

Download